Sign In to gain access to subscriptions and/or personal tools. |
|||||||||



Fromt the * Department of Orthopaedic Surgery, University of Connecticut Health Center, Farmington, Connecticut,
Orthopaedic Associates of Hartford, Hartford, Connecticut, and the
Department of Orthopaedic Surgery, Rush Medical College, Rush Presbyterian-St. Lukes Medical Center, Chicago, Illinois
Address correspondence to Robert A. Arciero, MD, Department of Orthopaedic Surgery, University of Connecticut Health Center, 263 Farmington Ave, Farmington, CT 06030 (e-mail: arciero{at}nso.uchc.edu).
Background: Recent technical improvements have led the way to a resurgence of the single-incision approach for repair of distal biceps tendon injuries. There has been no biomechanical evaluation of all these techniques with comparison to the standard 2-incision bone tunnel technique.
Hypothesis: There will be no difference under cyclic loading and ultimate failure between the 2-incision bone tunnel technique, suture anchor repair, interference screw, and EndoButton techniques for the repair of distal biceps tendon ruptures.
Study Design: Controlled laboratory study.
Methods: Sixty-three fresh-frozen cadaveric elbows were randomly assigned to 4 treatment groups (bone tunnel, EndoButton, suture anchor, interference screw). Cyclic loading was then performed from 0° to 90° at 0.5 Hz for 3600 cycles with a 50-N load. A differential variable reductance transducer was placed between the radius and distal tendon to determine displacement. The construct was then pulled to failure at 120 mm/min.
Results: A multiple analysis of variance revealed no statistically significant difference for displacement among the 4 repair techniques. Displacement using the bone tunnel was 3.55 mm, EndoButton was 3.42 mm, suture anchor was 2.33 mm, and interference screw was 2.15 mm. There was a statistically significant greater load to failure with EndoButton (440 N) than suture anchor (381 N), bone tunnel (310 N), or interference screw (232 N) (P < .001).
Conclusion: The EndoButton technique had the highest load to failure.
Clinical Relevance: These data demonstrate the EndoButton to be the strongest repair technique, with no failures during cycling at physiologic loads and with the largest load to failure. These findings are important in maximizing surgical results and stability and suggest that the construct can tolerate early postoperative active range of motion.
Key Words: biceps tendon rupture distal biceps tendon repair biomechanical evaluation
This article has been cited by other articles:
![]() |
J. King and M. Bollier Repair of Distal Biceps Tendon Ruptures Using the EndoButton J. Am. Acad. Ortho. Surg., August 1, 2008; 16(8): 490 - 494. [Full Text] [PDF] |
||||
![]() |
P. R. Chavan, T. R. Duquin, and L. J. Bisson Repair of the Ruptured Distal Biceps Tendon: A Systematic Review Am. J. Sports Med., August 1, 2008; 36(8): 1618 - 1624. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kettler, M. J. Tingart, J. Lunger, and V. Kuhn Reattachment of the distal tendon of biceps: FACTORS AFFECTING THE FAILURE STRENGTH OF THE REPAIR J Bone Joint Surg Br, January 1, 2008; 90-B(1): 103 - 106. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. Bisson, J. G. de Perio, A. E. Weber, M. T. Ehrensberger, and C. Buyea Is It Safe to Perform Aggressive Rehabilitation After Distal Biceps Tendon Repair Using the Modified 2-Incision Approach?: A Biomechanical Study Am. J. Sports Med., December 1, 2007; 35(12): 2045 - 2050. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | CONTACT US | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |