AJSM signin
HOME HELP CONTACT US SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     

Sign In to gain access to subscriptions and/or personal tools.
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Vangsness, C. T.
Right arrow Articles by Hedman, T. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Vangsness, C. T., Jr
Right arrow Articles by Hedman, T. P.
Related Collections
Right arrow Meniscus
Right arrow Laser/Radiofrequency energy
The American Journal of Sports Medicine 31:253-256 (2003)
© 2003 American Orthopaedic Society for Sports Medicine

Radiofrequency Thermal Effects on the Human Meniscus

An In Vitro Study of Systems with Monopolar and Bipolar Electrodes

C. Thomas Vangsness, Jr, MD*,, John D. Polousky, MD, Andrew B. Parkinson, MD and Thomas P. Hedman, PhD

From the Department of Orthopaedic Surgery, Keck School of Medicine, University of Southern California, Los Angeles, California

* Address correspondence and reprint requests to C. Thomas Vangsness, Jr., MD, University of Southern California Orthopaedic Surgery Associates, 1510 San Pablo Street, Suite 322, Los Angeles, CA 90033-4608

Background: No data exist on the cutting efficiency of monopolar versus bipolar radiofrequency energy application systems on human meniscal tissue.

Purpose: To compare the effects of monopolar and bipolar thermal energy systems on human meniscal tissue.

Study Design: Controlled laboratory study.

Methods: Fresh-frozen menisci were cut in cross-section into 180 pie-shaped specimens. A specially designed jig was used to consistently apply radiofrequency energy to the tissue under a constant 30-g force. Three different systems were tested at the low, middle, and high ranges, with application times of 1 and 3 seconds. Thermal effects were measured by image analysis microscopy.

Results: No significant differences in thermal effects were found with respect to energy output for each system. Both the individual system tested and the application time had statistically significant effects on thermal damage, with the individual system tested having a greater effect. The mean depths of thermal change produced by the Mitek (bipolar) device were 564 and 648 µm at 1 and 3 seconds applications, respectively. The Arthrocare device (bipolar) produced depths of 1444 and 1697 µm at 1 and 3 seconds. The Oratec device (monopolar) produced depths of 895 and 1057 µm, respectively.

Conclusions: A differential thermal effect was created in the meniscal tissue by three commercially available radiofrequency systems. Within the parameters of the experiment, all three systems limited thermal damage to a depth of less than 2 mm. The results appeared to depend more on the particular system used, not whether it had monopolar or bipolar electrodes.

Clinical Relevance: These data imply reasonably safe (less than 2 mm) thermal changes in the meniscus after radiofrequency energy application from these three systems.




This article has been cited by other articles:


Home page
JBJSHome page
P. M. Bonutti, T. M. Seyler, R. E. Delanois, M. McMahon, J. C. McCarthy, and M. A. Mont
Osteonecrosis of the Knee After Laser or Radiofrequency-Assisted Arthroscopy: Treatment with Minimally Invasive Knee Arthroplasty
J. Bone Joint Surg. Am., November 1, 2006; 88(suppl_3): 69 - 75.
[Abstract] [Full Text] [PDF]


Home page
Am J Sports MedHome page
A. Aneja, S. G. Karas, P. S. Weinhold, H. M. Afshari, and L. E. Dahners
Suture Plication, Thermal Shrinkage, and Sclerosing Agents: Effects on Rat Patellar Tendon Length and Biomechanical Strength
Am. J. Sports Med., November 1, 2005; 33(11): 1729 - 1734.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
M. D. Miller
What's New in Sports Medicine
J. Bone Joint Surg. Am., March 1, 2004; 86(3): 653 - 661.
[Full Text] [PDF]




HOME HELP CONTACT US SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2003 by the American Orthopaedic Society for Sports Medicine.